General Description

These two lines of balances have been selected to satisfy the requirements of either field or central testing laboratories. For more informat...
Read More

Description

These two lines of balances have been selected to satisfy the requirements of either field or central testing laboratories.

For more information concerning the selection criteria of balances see Extended information.
For the periodical checking, see Standard calibration masses in Related Products here below

Products

Code
Description
Units

Technical Specifications

First range: Supplied complete with Traceable Calibration Certificate

Models 11-

Capacity
[g]

Resolution
[g]

Pan dimensions
[mm]

Under balance
weighing facility

Voltage

Weigh
[kg] (approx.)

 
 

D0630/03A

300

0.001

dia. 80

No

110-230V/50-60Hz/1ph

1.5

 

D0630/05

520

0.001

dia. 80

Yes

110-230V/50-60Hz/1ph

1.5

 

D0630/06

600

0.01

dia. 130

Yes

110-230V/50-60Hz/1ph

1.5

 

D0630/3

3000

0.01

dia. 160

Yes

110-230V/50-60Hz/1ph

1.5

 

D0630/4A

4200

0.01

dia. 130

Yes

110-230V/50-60Hz/1ph

1.5

 

D0630/6

6000

0.1

225x300

Yes

110-230V/50-60Hz/1ph

3

 

D0630/15

15000

0.2

225x300

Yes

110-230V/50-60Hz/1ph

3

 

D0630/15A

15000

0.1

225x300

Yes

110-230V/50-60Hz/1ph

3

 

D0630/20

20000

0.1

175x195

Yes

110-230V/50-60Hz/1ph

4.6

 

D0630/30

30000

0.5

225x300

Yes

110-230V/50-60Hz/1ph

5.1

 

D0630/60A

60000

1

450x600

No

110-230V/50-60Hz/1ph

6.5

 

D0630/150A

150 [kg]

10

450x600

No

110-230V/50-60Hz/1ph

4

 

D0630/300

300 [kg]

20

450x600

No

110-230V/50-60Hz/1ph

4

 

 

Second range: DKD certificate (Deutscher Kalibrierdienst-German Accreditation Authority)

Models 11-

Capacity
[g]

Resolution
[g]

Pan dimensions
[mm]

Under balance
weighing facility

Voltage

Weigh
[kg] (approx.)

 
   

D0630/10

10000

0.1

150x170

Yes

230V/50-60Hz/1Ph

1.1

   

D0630/10Z

110V/60Hz/1Ph

   

D0630/24

24000

0.1

160x200

Yes

230V/50-60Hz/1Ph

4

   

D0630/24Z

110V/60Hz/1Ph

   

D0630/60

65000

1

340x240

No

230V/50-60Hz/1Ph

6.5

   

D0630/60Z

110V/60Hz/1Ph

   

D0632/150

60/150
[kg]

2 / 5

318x308

No

230V/50-60Hz/1Ph

4

   

D0632/150Z

110V/60Hz/1Ph

   

 

How to select balances

Where not specified by the relevant standards, the table given below may be used to select the appropriate balance conforming to the general specification for balances given in the ASTM D4753-95, EN 932-5 and BS 1377:1. Very often a single specification requires balances having a readability and tolerance of 0.1% of the specimen mass to be measured which is another suitable choice criteria. Other standards covering aggregates and granular materials in general give information about the quantity of aggregate to be tested depending upon its size. We have included this information in a separate table.

SELECTION GUIDE CONFORMING TO EN, ASTM, BS STANDARDS

Weighing ranges
(min capacity)

Standards

Readability
(scale interval
or digit)

Accuracy
(tolerance)

20 to 200 g

ASTM C114
(Cement)

0.2 mg

±0.2 mg

200 g

ASTM D4753(GP2)
EN 932-5
BS 1377:1

0.1 g
0.001 g
0.001 g

±0.2g
±0.005 g
±0.005 g

1200 g

ASTM D4753(GP2)
EN 932-5
BS 1377:1

0.2 g
0.001 g
0.001 g

±0.2g
±0.005 g
±0.005 g

2000 g

ASTM D4753(GP2)
EN 932-5
BS 1377:1

1 g
0.1 g
0.1 g

±2g
±0.3 g
±0.3 g

5000 g

ASTM D4753(GP2)
EN 932-5
BS 1377:1

2 g
0.2 g
0.1 g

±2g
±0.3 g
±0.3 g

10000 g

ASTM D4753(GP2)
EN 932-5
BS 1377:1

5 g
1 g
1 g

±5g
±3 g
±3 g

25000 g

ASTM D4753(GP2)
EN 932-5
BS 1377:1

5 g
5 g
5 g

±5 g
±10 g
±10 g

50000 g

ASTM D4753(GP2)
EN 932-5
BS 1377:1

50 g
10 g
10 g

±5 0g
±30 g
±20 g


 

 

MINIMUM MASS OF SAMPLE FOR SIEVE ANALYSIS (bs 812:1)

Nominal size of material
mm

Minimum mass of sample
kg

63

50

50

35

40

15

28

5

20

2

14

1

10

0.5

6

0.2

5

0.2

3

0.2

<3

0.1

 

 

Standards

  • EN 932-5
  • ASTM D4753